User guide
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 554/754 pump is intended for use in the electromagnetic environment specified below.
The customer or user of the Paradigm 554/754 pump system should make sure that it is used in such
an environment.
Immunity
Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Paradigm
554/754 pump, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
10 Vrms
150 kHz to
80 MHz
Not applicable Not applicable
Pump specifications 221
Chapter 15