Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 554/754 pump system (consisting of the MMT-554/754 pump and MMT-7703 transmitter)
is intended for use in the electromagnetic environment specified below. The customer or the user of
the Paradigm 554/754 pump system should make sure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±30 kV air
(<5% relative
humidity)
The Paradigm 554/754 pump
system should not be affected
by electrostatic discharge that
might occur under normal
conditions of use.
Electrical fast transient/
burst
±2 kV for power supply
lines
Not applicable Not applicable
IEC 61000-4-4
±1 kV for input/output
lines
Not applicable Not applicable
Surge
±1 kV line(s) to line(s) Not applicable Not applicable
IEC 61000-4-5
±2 kV line(s) to earth Not applicable Not applicable
Voltage dips, short
interruptions and voltage
variations on power
supply lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
Not applicable Not applicable
40% U
T
(60% dip in U
T
)
for 5 cycles
Not applicable Not applicable
70% U
T
(30% dip in U
T
)
for 25 cycles
Not applicable Not applicable
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Not applicable Not applicable
Power frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Pump specifications220