User guide

Guidance and manufacturer's declaration
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Paradigm 554/754 pump system (consisting of the MMT-554/754 pump and MMT-7703 transmitter)
is intended for use in the electromagnetic environment specified below. The customer or the user of
the Paradigm 554/754 pump system should make sure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The Paradigm 554/754 pump system uses RF energy
only for system communication functions.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B The Paradigm 554/754 pump system is suitable for
use in all establishments other than domestic and
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
NOTE: The preceding statement is required by
IEC 60601-1-2 for Group 1, Class B devices.
However, since the Paradigm 554/754 pump system
is battery powered, its emissions will not be
affected by the establishment power supply and
there is no evidence of any issues associated with
the use of the system in domestic establishments.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Pump specifications 219