User guide
Feature Options Default Setting Limits, Range of Values
Alert Silence • Off
• Hi Alerts
• Lo Alerts
• Hi and Lo Alerts
• All Sensor Alerts
Off If any option other than Off
is selected in the ALERT
SILENCE screen, the SET
DURATION screen displays.
The value range is from
00:30 to 24:00.
AUC Limits Always on Low Limit
3.9 mmol/L
(70 mg/dL)
High Limit
10 mmol/L
(180 mg/dL)
Low Limit range:
d
2.2–22.2 mmol/L (40–
400 mg/dL)
High Limit range:
e
2.2–22.2 mmol/L (40–
400 mg/dL)
a. Your Low Glucose Limit must be at least 0.6 mmol/L (10 mg/dL) below your High Glucose Limit. For example, if your High
Glucose limit is 11.1 mmol/L (200 mg/dL), then your Low Glucose limit cannot be higher than 10.5 mmol/L (190 mg/dL).
b. Your High Glucose Limit must be at least 0.6 mmol/L (10 mg/dL) above your Low Glucose Limit.
c. If you select this option the sensor glucose graphs will show continuously on the pump screen.
d. The AUC Low Limit can be equal to, but not higher than the AUC High Limit.
e. The AUC High Limit can be equal to, but not lower than the AUC Low Limit.
Pump specifications218