User guide
Feature Options Default Setting Limits, Range of Values
Rate of Change
Alerts
• On
• Off
Fall Rate Limit
Off
Rise Rate Limit
Off
Fall Rate Limit range:
0.065–0.275 mmol/L/min
(1.1–5.0 mg/dL/min)
Rise Rate Limit range:
0.065–0.275 mmol/L/min
(1.1–5.0 mg/dL/min)
Low Glucose
Suspend
• On
• Off
Off Range:
3.3–6.1 mmol/L (60–
110 mg/dL)
Cal Repeat (Always on - cannot be turned
off)
On 30 minutes Range:
0:05 – 1:00 (5 minutes to
1 hour)
Cal Reminder • On
• Off
On 1 hour Range:
0:05 – 6:00 (5 minutes to
6 hours)
Auto Calibration • On
• Off
Off —
Blood Glucose
Units (BG Units)
• mg/dL
• mmol/L
mg/dL —
Weak Signal (Always on - cannot be turned
off)
30 minutes Range:
0:05 – 0:40 (5 to 40 minutes)
Sensor glucose
graph timeout
• 2 minutes
• 4 minutes
• 6 minutes
• NONE
c
2 minutes —
Pump specifications 217
Chapter 15