User guide
Bolus Wizard feature examples
Settings:
• Carb ratio: 30 grams/unit
• BG Target: 5.0-6.7 mmol/L (90-120 mg/dL)
• Insulin Sensitivity: 2.2 mmol/L/units (40 mg/dL/unit)
• Active Insulin Time: 6 Hours
1 No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter BG.
Food estimate:
60 g
30 g/u
= 2 units
2 units
Bolus estimate:
2 No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a BG of 11.1 mmol/L
(200 mg/dL).
Food estimate:
60 g
30 g/u
= 2 units
= 2 units
Correction estimate:
11.1 mmol/L - 6.7 mmol/L
2.2 mmol/L/u
= 2 units
200 mg/dL - 120 mg/dL
40 mg/dL/u
Bolus estimate = Food estimate + Correction estimate
Bolus estimate = 2 units + 2 units = 4 units
or
Pump specifications210