User guide
Square Wave bolus
Delivers bolus insulin over a duration of 30 minutes up to eight hours (limited by the Max bolus setting).
Temporary (temp) basal rate
Allows you to temporarily change the current basal rate for a duration of 30 minutes up to 24 hours (limited
by Maximum basal setting). The temp basal rate can be set to either Percent of basal or Insulin rate.
Time and date screen
12-hour or 24-hour formats. Pump users set the time/date, including the year, month and day. The date is
included in the STATUS screen. Time always appears on the top of the screen.
Environmental conditions
• Pump operating temperature range: from 3°C (37°F) to 40°C (104°F)
• Air pressure range: from 700 hPa to 1060 hPa (from 10.2 psi to 15.4 psi)
• Operating humidity range of the case: 20% to 95%. This requirement exceeds EN 60601-2-24:1998, Section
10.2.1b (20% to 90%).
Status screen
Item When What
Active Insulin (if any) (Active Ins:) the amount of active
insulin, displayed to the next 0.025 U
Alarm Clock* (appears if scheduled to
alarm)
Time set is displayed.
Alert Silence (if active) Hi, Lo, Hi/Lo, All with time left
Auto-off (appears if enabled) X HR
Basal pattern information (if active) Pattern A or B
Battery Status (always appears) Normal, Low, Off
BG meter value
(most recent BG value
received)
(appears if BG meter is
enabled)
XX.X mmol/L (XXX mg/dL) time and date
received
Pump specifications204