User guide

Easy bolus
Features programming using audible tones (or vibrate pulses) in user-defined increments.
Beep mode range: 0 to maximum bolus
Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first
Factory default: off
Default step increment: 0.1 unit
Step size maximum bolus.
Adjustable step size: 0.1 to 2.0 units per step.
Accessible from the remote control or pump buttons.
Filling the infusion set
Filling the cannula: up to 10 units (limited by maximum bolus)
Filling the tubing: warning at 30 units, then at each 10 units thereafter.
Fill rate: 1 to 5 units/second.
Insulin used to fill the infusion set is not counted in daily totals but is recorded separately in the Reservoir
+ Set history.
Infusion pressure
Maximum infusion pressure and occlusion pressure: 13.7 PSI (94.46 kPa).
Insulin sensitivity
Maximum settings: 8
Factory default: 2.8 mmol/L/u (50 mg/dL/u)
Range: 0.5–22.2 mmol/L/u (10–400 mg/dL/u)
Warning limits: less than 1.1 or greater than 5.6 mmol/L/u (less than 20 or greater than 100 mg/dL/u)
Pump specifications 201
Chapter 15