User guide
Bolus units
Increments: Based on selected scroll rate
Bolus Wizard feature
See the Bolus Wizard feature specifications section in this chapter for information.
Carb ratios
Maximum ratio settings Range Warning limits
8
1–200 grams/u
less than 5 or greater than 50 grams/u
0.075–
15.0 u/exch
less than 0.3 or greater than 3.0 u/exch
Carb units
The food entry when using Bolus Wizard feature:
• grams: 0–300 (increments: 1 gram)
• exchanges: 0.0–20 (increments: 0.5 exch)
Daily totals
Maximum records shown: 32 days of data, maximum display: 999.975 units/day. Daily delivery accuracy:
0.025 units.
Default screen
The HOME screen. For most screens, if no buttons are pressed for 30 seconds, the pump returns to this screen.
Delivery accuracy
• Delivery accuracy: +/-5 percent.
• All normal boluses will be delivered within a maximum of 28 minutes.
• Displacement (piston) accuracy: +/-2 percent.
Pump specifications 199
Chapter 15