User guide

Backlight
LCD (Liquid Crystal Display) type
Time-out: 30 seconds
Basal
Delivery: 0.025–35 units/hour (maximum units: 35/hour)
Factory maximum setting: 2.0 units/hour
3 patterns maximum, each with 48 rates
Increments:
0.025 units for basal amounts in the range of 0.025 to 0.975 units
0.05 units for basal amounts in the range of 1 to 9.95 units
0.1 units for basal amounts of 10.0 units or larger
BG Target
Maximum targets: 8
Range: 3.3–13.9 mmol/L (60–250 mg/dL)
Warning limits: less than 5.0 or greater than 7.8 mmol/L (less than 90 or greater than 140 mg/dL)
Bolus delivery
Insulin delivered/stroke:
0.025 units for bolus amounts in the range of 0.025 to 0.975 units
0.05 units for bolus amounts larger than 0.975 units
Fluid delivered/stroke: 0.25 µL (microliter) for 0.025 unit pump stroke; 0.5 µL for 0.05 unit pump stroke
Maximum bolus
Range: 0.0–75.0 units
Warning limit: greater than 25.0 units
Bolus history
Maximum records shown: 24
Pump specifications198