User guide
Backlight
• LCD (Liquid Crystal Display) type
• Time-out: 30 seconds
Basal
• Delivery: 0.025–35 units/hour (maximum units: 35/hour)
• Factory maximum setting: 2.0 units/hour
• 3 patterns maximum, each with 48 rates
• Increments:
• 0.025 units for basal amounts in the range of 0.025 to 0.975 units
• 0.05 units for basal amounts in the range of 1 to 9.95 units
• 0.1 units for basal amounts of 10.0 units or larger
BG Target
Maximum targets: 8
• Range: 3.3–13.9 mmol/L (60–250 mg/dL)
• Warning limits: less than 5.0 or greater than 7.8 mmol/L (less than 90 or greater than 140 mg/dL)
Bolus delivery
• Insulin delivered/stroke:
• 0.025 units for bolus amounts in the range of 0.025 to 0.975 units
• 0.05 units for bolus amounts larger than 0.975 units
• Fluid delivered/stroke: 0.25 µL (microliter) for 0.025 unit pump stroke; 0.5 µL for 0.05 unit pump stroke
Maximum bolus
• Range: 0.0–75.0 units
• Warning limit: greater than 25.0 units
Bolus history
Maximum records shown: 24
Pump specifications198