User guide
Troubleshooting sensor features
Reconnect old sensor
You should only use this feature if you have disconnected the sensor from the transmitter and have to
reconnect them.
To Reconnect old sensor:
1 Go to the RECONNECT OLD SENSOR screen.
Main > Sensor > Link to Sensor > Reconnect Old Sensor
2 Press ACT in the RECONNECT OLD SENSOR screen.
3 The SENSOR READY 2 HRS screen appears. Press any button to continue. An alert will sound in two hours
when a meter BG is required for calibration.
Find lost sensor
If you receive a Lost Sensor alert:
1 Relocate the pump closer to the sensor, and go to the LINK SENSOR MENU.
Main > Sensor > Link to Sensor
2 Select Find Lost Sensor, then press ACT.
3 The SENSOR READY 15 MIN screen appears. Press any button to continue. An alert will sound in 15 minutes
when a meter BG is required for calibration.
Understanding your transmitter, tester, and charger
The charger has a green light that shows the charging status and a red light that communicates any problems
during charging. See the following graphic for the location of these lights:
Question: Why did the flashing green charger
light turn off and a longer flashing red
charger light turn on during charging?
about 2 secs about 2 secs
Troubleshooting, alerts and alarms 191
Chapter 13