User guide

NOTE: Two hours after you choose to keep your insulin delivery suspended, the pump automatically
resumes delivering insulin at the basal rate (standard, pattern or temporary) that is programmed to be
active for that time. The pump returns to the HOME screen as insulin delivery resumes. When you press
ACT from the HOME screen, the following message appears: Basal delivery is resumed. Check BG.
Press any key to cont.
To resume insulin delivery after you suspend your pump:
1 Press ACT from the HOME screen when you are ready to restart your basal delivery.
2 The LOW SUSPEND/CHECK BG screen appears. Select Restart Basal, then press ACT.
3 The pump resumes delivering insulin at the basal rate (standard, pattern or temporary) that is
programmed to be active for that time. The following message appears as insulin delivery resumes: Basal
delivery is resumed. Check BG. Press any key to cont.
4 Press any button to continue using your pump.
Testing the transmitter
The tester acts like a sensor. If you get sensor-related alerts, use the tester to make sure that the transmitter
is working.
Connecting the tester
1 Hold the transmitter and the tester as shown. Line up the flat side of the tester with the flat side of the
transmitter.
2 Push the tester into the transmitter until the flexible side arms of the tester click into the notches on
both sides of the transmitter.
Troubleshooting, alerts and alarms 189
Chapter 13