User guide

MAX DELIVERY
This alarm alerts you when you have taken more insulin than expected based on Maximum bolus and Maximum
basal rates.
MOTOR ERROR
Insulin delivery has stopped. This alarm occurs if your pump detects a motor error.
NO DELIVERY
Insulin delivery has stopped. This alarm will occur if your pump detects a blockage or if the reservoir is empty.
Refer to the first page of the Troubleshooting chapter for additional information.
NO RESERVOIR
The reservoir is not inserted correctly or no reservoir has been inserted.
OFF NO POWER
The battery is dead. Replace battery immediately. Follow the directions on the screen. Check to make sure
that the time is correct on the screen. Reset the time if necessary.
RESET
The Reset alarm triggers when pump settings are cleared because of one of these reasons:
Pump settings were cleared (Clear Settings function) and settings have not been reprogrammed.
A download attempt from the PC is incomplete. (The download function is applicable to the optional
software feature. Refer to the software user guide for more information.)
WEAK BATTERY
The pump tests the voltage of each battery installed. If the battery voltage is less than full strength, this
alarm may occur. The pump will operate normally, but the battery life will be shorter than expected. Always
make sure to install a new battery in the pump.
Troubleshooting, alerts and alarms 187
Chapter 13