User guide

CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/date.
E (ERROR)
An E-error alarm will show the letter E followed by two numbers. E-alarms cause all insulin delivery to stop,
the pump resets, and all your settings are cleared. After receiving this alarm, note the error number then
call your local Medtronic Diabetes representative for assistance.
EMPTY RESERVOIR
There is no insulin in the reservoir. Change the reservoir immediately.
FAILED BATT TEST
The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not used.
If the battery does not have enough voltage, this alarm will occur. The pump will not function and the battery
must be replaced. (Always make sure that you install a NEW battery into the pump.)
FINISH LOADING
You have not completed filling the infusion set with insulin. Clear the alarm. This resumes basal delivery.
See the Filling the cannula section in the Starting on insulin chapter to fill the infusion set cannula with
insulin. If your infusion set has a needle cannula, press ESC to skip this step.
MAX FILL REACHED
If you use more than 30 units of insulin to fill the infusion set tubing, this message appears. Press ESC, ACT
to clear the message. See the Filling the tubing section in the Starting on insulin chapter for more
information.
Troubleshooting, alerts and alarms186