HIGH PREDICTED
Reason: The sensor glucose measurements may reach or go above your High Glucose Limit in the length of
time you selected for the High Predictive alert. The pump plays three consecutive tones, in rising pitch, if
an audible beep has been selected as the alert type.
RISE RATE
Reason: The sensor glucose measurements are rising at a rate that is equal to or faster than the Set Rise
Rate Limit you selected for the alert. The pump plays two consecutive tones, in rising pitch, if an audible
beep has been selected as the alert type.
FALL RATE
Reason: The sensor glucose measurements are falling at a rate that is equal to or faster than the Set Fall
Rate Limit you selected for the alert. The pump plays two consecutive tones, in falling pitch, if an audible
beep has been selected as the alert type.
SENSOR ERROR
Reason: The sensor signals are either too high or too low.
Action: You do not need to change the sensor. Clear the alert. If the alert persists, test your transmitter
with the tester.
Alarms
Your pump has a sophisticated network of safety checks and systems. If the safety network detects anything
unusual, your pump notifies you of conditions that require your immediate attention. The backlight
illuminates the pump screen and the alarm message displays on the screen.
It is important that you understand how your pump behaves when you do not clear an alarm or an alert. See
the Alerts section in this chapter for an explanation and an example.
Check your pump and test your blood sugar. Follow the guidelines prescribed by your healthcare professional
to manage your diabetes as needed.