User guide

Action: Follow the guidelines in the Calibrating the sensor section in the Sensor features chapter to enter
a new meter BG for calibration. If you receive a Cal Error on your second calibration, a CHANGE SENSOR alert
will occur. Call your local representative if you have questions.
METER BG NOW
Reason: A meter BG measurement is needed immediately to calibrate your sensor so that you can continue
receiving sensor glucose readings.
Action: Take and enter a meter blood glucose measurement. Follow the guidelines in the Calibrating the
sensor section in the Sensor features chapter.
METER BG BY
Reason: A meter BG measurement must be entered by the time that is shown to calibrate your sensor. The
METER BG BY alert is also known as the CAL REMINDER alert.
Action: Enter a blood glucose measurement to avoid a METER BG NOW alert.
LOW X.X MMOL/L (XX MG/DL)
Reason: The sensor glucose number is equal to or lower than your low glucose limit. If you do not set a Low
Glucose Limit and do not set the glucose alerts ON, then you will not get a Low sensor glucose alert. The
pump plays four consecutive tones, in falling pitch, if an audible beep has been selected as the alert type.
HIGH XX.X MMOL/L (XXX MG/DL)
Reason: The sensor glucose number is higher than or equal to your high glucose limit. If you do not set a High
Glucose Limit and do not set the glucose alerts ON, then you will not get a High sensor glucose alert. The
pump plays four consecutive tones, in rising pitch, if an audible beep has been selected as the alert type.
LOW PREDICTED
Reason: The sensor glucose measurements may reach or go below your Low Glucose Limit in the length of
time you selected for the Low Predictive alert. The pump plays three consecutive tones, in falling pitch, if
an audible beep has been selected as the alert type.
Troubleshooting, alerts and alarms182