User guide
What is a CHECK SETTINGS alarm?
This alarm occurs after an E alarm or after you clear your pump. It is advising you to make sure that all your
settings are correct. A CHECK SETTINGS alarm occurs after any of these actions:
• All user settings were cleared (set back to their defaults) because there was an E-error alarm,
• The Clear Settings function was performed, or
• After you rewind when practicing without a reservoir when you first get your pump. In this case, it is
just a reminder to make sure all your settings are programmed before you begin using the pump with
insulin.
My screen appears distorted
The screen may appear distorted or have a rainbow-like appearance if you are wearing polarized sunglasses,
are in bright sunlight, or in extreme high or low temperatures. If your screen appears distorted:
• Take off your sunglasses.
• Move into the shade.
• Make sure your pump is not in direct heat (next to a heater) or cold (worn on the outside of your clothing
on a very cold day).
• Do not return the pump: this is a normal property of this type of screen on any device.
The pump is asking me to rewind
This is normal after any of the following:
1 Any E-alarms,
2 The Clear Settings function,
3 Or, a NO DELIVERY alarm (during the Reservoir + Set sequence).
My bolus stopped
The Bolus Stopped error can occur if the battery cap is loose or the pump was bumped or dropped during a
bolus. It can also happen if the pump receives a static shock. As a safety measure, the pump stops the bolus
when this happens.
1 If you dropped your pump, visually inspect it to make sure that it is not damaged in any way.
2 Review your bolus history and reprogram the remaining bolus, if needed.
Troubleshooting, alerts and alarms 175
Chapter 13