User guide

Auto-off
The factory setting for this feature is Off. This is a safety feature that stops insulin delivery after a defined
time period (from 1 to 24 hours). If the pump detects that no buttons have been pressed for the selected
amount of time in Auto-off, insulin delivery will stop and an alarm will sound. You may choose to program
this feature into your pump based on the number of hours that you usually sleep at night. Discuss what uses
and settings are best for you with your healthcare professional.
1 Go to the AUTO OFF DURATION screen.
Main > Utilities > Alarm > Auto Off
2 Set the number of hours you want to set and press ACT.
If you do not want to use the Auto Off feature, make sure the hour is set to zero (0).
3 The screen will return to the ALARM MENU. The Auto Off feature is now set. Exit the menus.
LOW RESV WARNING (Low reservoir warning)
Allows you to program the pump to sound an alert before your reservoir is empty. You can select one of these
warning types:
a specified number of units that remain in the reservoir
a specified maximum amount of time that remains before the reservoir will be empty
The factory setting for this feature is 20 insulin units.
If you use Time as the low reservoir warning type and you deliver large boluses, the actual time remaining
could be less than the warning time. The warning time is based on the basal insulin delivery rate. The Time
low reservoir warning type is intended to let you know if you will have enough insulin while you are sleeping.
1 Go to the RESV WARNING TYPE screen.
Main > Utilities > Alarm > Low Resv Warning
2 Select Insulin Units or Time and press ACT.
WARNING: When the pump detects a low reservoir condition during a bolus or Fill Cannula delivery,
the LOW RESERVOIR alert displays after the delivery is finished. Make sure to check the volume of
your reservoir to ensure enough insulin is available.
a. For Insulin units:
In the WARNING UNITS screen enter the number of units you want remaining for the first warning,
then press ACT.
Utilities154