and the pump beeps three times. Go to the ENTER BG screen of the Bolus Wizard feature to
enter a correction bolus. See the Using the Bolus Wizard feature chapter for more information.
If the information is not correct, select Cancel, then press ACT. The SENSOR MENU screen
reappears with Calibrate selected. Repeat the procedure and enter the correct information.
c. To enter a meter BG through the Capture Event Menu, follow this path:
Main > Capture Event > Enter BG
See the Entering BG measurements section in the Basic programming chapter for more information.
d. If at the time of sensor calibration you also need to perform a bolus using the Bolus Wizard, you can
use the meter BG for both calibrating the system and bolusing. To enter a meter BG value manually
in the Bolus Wizard, press the
button on your pump, or use the MAIN MENU to go to the ENTER BG
screen:
Main > Bolus > Use Bolus Wizard
2 If you are manually entering your meter BG, use the up and down arrow buttons on your pump to do so.
Press ACT. For the Bolus Wizard option above, select Yes when the following screen displays: BG to
update Sensor?
Sensor features 139