User guide
Setting the Cal Reminder
The Cal Reminder feature allows you to set a reminder to calibrate your system. For example, if you set your
reminder to four hours, then you will receive a METER BG BY alert (Cal Reminder) four hours before the next
meter blood glucose (BG) entry is due (eight hours after your last successful sensor calibration).
To set up the Cal Reminder:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Cal Reminder, then press ACT. The CAL REMINDER screen shows On.
3 Press ACT. The SET CAL REMINDER screen flashes the Cal Reminder default time of 1 hour (1:00).
4 Select your Cal Reminder time between 5 minutes (0:05) and 6 hours (6:00).
5 Press ACT. The EDIT SETTINGS screen shows the Cal Reminder time you selected. You are now ready to
set up the Auto Calibrate feature.
Turning on the Auto Calibration
Auto Calibration feature determines how to calibrate the pump when using a linked meter. When this feature
is turned on, the pump uses every linked meter BG reading in the range of 2.2–22.2 mmol/L (40–400 mg/dL)
for calibration. When this feature is turned off, the pump asks you every time if you want to use the linked
meter BG for calibration.
To turn on the Auto Calibration:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Auto Calibrate, then press ACT. The AUTO CALIBRATION screen is displayed.
3 Select On, then press ACT. The EDIT SETTINGS screen appears showing that the Auto Calibration is now
turned On. You are now ready to enter the ID number of your transmitter.
Entering the transmitter ID
The transmitter ID (serial number) starts with 2 and is found on the flat side of your transmitter. You must
enter the transmitter ID so that the transmitter and the pump can communicate with each other.
Sensor features126