Setting the Low Repeat
After you receive and clear a Low Glucose, Fall Rate of Change, or Low Predictive alert, the alert will repeat
until the condition that caused the alert is resolved. After you clear a Low Suspend alarm and resume basal
delivery, the alarm will repeat if a low glucose condition still exists. The Low Repeat feature allows you to
set how frequently you want the above alerts or alarm to repeat after you clear it the first time. The Low
Repeat feature works similar to the High Repeat one. See Setting the High Repeat section in this chapter for
an example.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Low Repeat, then press ACT. The SET LO ALERT REPEAT screen flashes the default Lo Alert Repeat
time of 20 minutes (0:20).
3 Select your Low Repeat time. The time must be between 5 minutes (0:05) and 1 hour (1:00).
4 Press ACT. The EDIT SETTINGS screen shows the Low Repeat value set. You are now ready to set up the
Predictive Alert.
Setting the Predictive glucose alert
The Predictive alerts calculate when you may reach your Low or High Glucose Limits, then send you an alert
before you reach those limits. A Predictive alert tells you that if your sensor glucose measurements keep
falling or rising at the current rate, you may reach your Glucose Limit in the number of minutes you selected.
For a LOW PREDICTED alert, the pump plays three consecutive tones, in falling pitch, if an audible beep has
been selected as the alert type. For a HIGH PREDICTED alert, the pump plays three consecutive tones, in
rising pitch, if an audible beep has been selected as the alert type.