User guide

6 To set a second pair of Glucose Limits, select the time of day you want this pair to become active.
7 Press ACT. The SET GLUCOSE LIMITS 2 screen appears. The default Low Glucose Limit flashes (OFF).
8 Select your Low Glucose Limit. The value must be between 2.2 and 21.6 mmol/L (40 and 390 mg/dL).
9 Press ACT. The default High Glucose Limit flashes (OFF).
10 Select your High Glucose Limit, then press ACT. The value must be between 2.8 and 22.2 mmol/L (50
and 400 mg/dL).
11 Press ACT. The SET START TIME 3 screen appears.
The second pair of Glucose Limits (Glucose Limits 2) is now set. If you do not need any more Glucose
Limits pairs, press ESC. If you want to set more Glucose Limits, then repeat steps 6 through 11 above
for each pair, up to a maximum of eight.
You are ready to set up your High Repeat.
Setting the High Repeat
After you receive and clear a High Glucose, Rise Rate of Change, or High Predictive alert, the alert will repeat
until the condition that has caused the alert is resolved. The High Repeat feature allows you to set how
frequently you want the alert to repeat after you clear it the first time. For example, if your blood glucose
levels usually take one to two hours to fall after a correction bolus, you may want to set the High Repeat to
one or two hours to avoid unnecessary alerts.
Example
You set your High Repeat to 15 minutes. You receive an alert at 13:00 and immediately clear it. However,
the condition that caused the alert still exists, so you will receive a second alert at 13:15, 15 minutes after
you cleared the first alert. If you immediately clear the alert again, then the alert will repeat at 13:30 and
continue to repeat every 15 minutes until you resolve the condition that caused the alert. (This example also
applies to the Low Repeat.)
To set up your High Repeat:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Hi Repeat, then press ACT. The SET HI ALERT REPEAT screen flashes the default High Repeat time
of 1 hour (1:00).
3 Select your High Repeat time. The time must be between 5 minutes (0:05) and 3 hours (3:00).
4 Press ACT. The EDIT SETTINGS screen shows the High Repeat value set.
You are now ready to set up your Low Repeat.
Sensor features 121
Chapter 8