User guide

Setting the Glucose Limits start time
If your healthcare professional recommends that you set more than one pair of Glucose Limits, you must
enter a start time for each. For example, you may use one pair of Glucose Limits during the day and another
pair at night. The example below shows two pairs of Glucose Limits. The first pair begins at midnight, and
the second pair begins at 9:00.
00:00 09:00
00:00
Glucose Limits 1
High Glucose Limit:
13.3 mmol/L (240 mg/dL)
Low Glucose Limit:
3.3 mmol/L (60 mg/dL)
00:00 - 09:00
(9 hours)
Glucose Limits 2
High Glucose Limit:
13.9 mmol/L (250 mg/dL)
Low Glucose Limit:
4.4 mmol/L (80 mg/dL)
09:00 - 00:00
(15 hours)
To set up your Glucose Limits:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Glucose Limits, then press ACT. The SET GLUCOSE LIMITS 1 screen appears. The Low Glucose
Limit flashes 4.4 mmol/L (80 mg/dL).
The start time for your first pair of Glucose Limits is midnight and cannot be changed.
3 Select your Low Glucose Limit. The value must be between 2.2 and 21.6 mmol/L (40 and 390 mg/dL).
To switch off the Low Glucose Limit, reduce it below 2.2 mmol/L (40 mg/dL) to select OFF.
4 Press ACT to select your High Glucose Limit. The screen flashes 13.3 mmol/L (240 mg/dL). The limit you
select must be between 2.8 and 22.2 mmol/L (50 and 400 mg/dL). Your High Glucose Limit must also be
at least 0.6 mmol/L (10 mg/dL) above your Low Glucose Limit. To switch off the High Glucose Limit,
reduce the High Glucose Limit below 2.8 mmol/L (50 mg/dL) to select OFF.
5 Press ACT. The SET START TIME 2 screen appears. The dashes under the screen name flash. The first pair
of Glucose Limits is now set.
If you do not need a second pair of Glucose Limits, press ESC and skip to the next section. If you want
to set up a second pair of Glucose Limits, follow steps 6 through 11 below.
Sensor features120