User guide
Turning on the Glucose Alerts
The Glucose alerts must be turned on if you want the system to send you an alert when the sensor glucose
measurements reach or exceed your Glucose Limits. For a Low sensor glucose alert, the pump plays four
consecutive tones, in falling pitch, if an audible beep has been selected as the alert type. For a High sensor
glucose alert, the pump plays four consecutive tones, in rising pitch, if an audible beep has been selected
as the alert type.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Glucose Alerts, then press ACT. The GLUCOSE ALERTS screen appears.
3 Select On, then press ACT. The EDIT SETTINGS screen appears showing the glucose alerts are now turned
On.
Setting the Glucose Limits
If you turned the Glucose alerts On, then you need to set the high and low Glucose Limits recommended by
your healthcare professional. Your recommended glucose limits may vary throughout the day, so your pump
allows you to set up to eight pairs.
A Glucose Limit pair includes one High Glucose Limit and one Low Glucose Limit, as shown in the following
table:
Glucose Limit pair
Alerts
Low: 4.4 mmol/L (80 mg/dL) A Low Glucose Limit alert occurs when the sensor glucose
measurement reaches or goes below the Low Glucose Limit in this
table.
High: 13.3 mmol/L (240 mg/dL) A High Glucose Limit alert occurs when the sensor glucose
measurement reaches or goes above the High Glucose Limit in this
table.
Sensor features118