User guide
You cannot make changes to your normal basal rate while a percent temp basal is active. You must either
wait until the temp basal is finished or cancel the temp basal in order to reprogram your normal basal rate
setting(s).
1.50 U/H
(rate 1)
2.55 U/H
(rate 2)
1.25
U/H
1.95 U/H
(rate 3)
(50% x rate 3 = 0.975)
0.975 rounded down to
the next 0.05 increment
is 0.95 U/H
(50% x rate 2 = 1.275)
1.275 rounded down to
the next 0.05 increment is
1.25 U/H
07:00
(7:00 a.m.)
13:05
(1:05 p.m.)
15:00
(3:00 p.m.)
17:05
(5:05 p.m.)
temp basal stop time
00:00
(12:00 a.m.)
Temp Basal Settings
Temp basal type: Percent of basal
Duration: 4 hours (13:05-17:05)
Rate: 50%
temp basal start time
0.95
U/H
Optimizing pump therapy 111