Turning on the patterns
Your pump is set at the factory with the basal patterns feature turned off. After you turn on patterns, you
still have to program and select a pattern (A or B), as described in the next sections, before the patterns
feature is active. If you turn off the patterns feature, your pump will automatically select your standard
basal pattern.
1 Go to the PATTERNS OPTION screen.
Main > Basal > Basal Setup > Patterns
2 Select On, then press ACT. The patterns feature is now on. Exit the menus.
Programming a pattern
Your pump will keep your pattern settings even when the Patterns option is turned off. However, the patterns
feature must be on to program a basal pattern.
To program your patterns, do these steps:
1 Go to the EDIT BASAL screen.
Main > Basal > Basal Setup > Set/Edit Basal
2 Select the basal pattern you want to program, then press ACT.
3 The SET BASAL RATE 1 screen appears. The basal rate flashes, indicating that it can be changed. Set your
first rate, then press ACT.
The start time for your first basal rate is midnight and cannot be changed.
4 The SET START TIME 2 screen appears. The dashes under the screen name flash. The first basal rate is
now set.
If you want to use the same rate for the whole day, press ESC and skip to step 10. If you want to program
another basal rate, follow steps 5 through 8.
5 In the SET START TIME 2 screen enter the time of the day you want this basal rate to become active.
6 Press ACT. The SET BASAL RATE 2 screen appears.
7 The previously set basal rate or dashes flash on the screen. Select the value for this basal rate.
8 Press ACT. The SET START TIME 3 screen appears. The second basal rate is now set.
Optimizing pump therapy106