User guide

Step value setup
You can set the step value from 0.1 to 2.0 units, or to the set maximum bolus limit of less than 2.0 units
(factory setting: 0.1). Set the step to a number that is convenient to use and easy to multiply.
Example:
2.0 unit Easy bolus with a step
amount set at 0.5 units
Total # Steps = 4
Total # Button Presses = 4
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
4
3
2
1
0
0 0.5 1.0 1.5 2
Total # Units
1 The EASY BOLUS ENTRY screen flashes the step value. Change the value, then press ACT. The step value
is the increment you will use for your Easy bolus.
2 The screen will return to the BOLUS MENU. Your step amount is now programmed and Easy Bolus is ready
to use. Exit the menus.
Delivering Easy Bolus
Practice using the Easy Bolus feature while looking at the pump screen as you count the beeps. This feature
only works from the HOME screen. After you are familiar with Easy Bolus, you can use the audible tones for
bolus delivery without having to look at the screen.
1
From the HOME screen, press
. The SET EASY BOLUS screen appears with the single step value flashing.
2
Press
the number of times needed for your bolus amount. Watch the amount change on the screen
with each press. The pump will vibrate or sound a different tone for each
press.
Optimizing pump therapy102