User guide

7 The SQUARE DURATION screen appears. Enter the amount of time you want the Square Wave bolus to
last, then press ACT.
NOTE: If you have BG Reminder turned On, the BG REMINDER DURATION screen displays. It allows you
to set the duration before you are reminded to check your blood glucose after a bolus. See the BG
Reminder section in the Basic programming chapter for information about this feature.
8 Press ACT to accept and deliver the bolus. The BOLUS DELIVERY screen appears with an open circle
indicating that your pump is in Special Mode. The pump beeps or vibrates at the start of the bolus. During
bolus delivery, the pump returns to the HOME screen. If you want to see the progress of the delivery, press
ESC to see the STATUS screen. The pump beeps or vibrates at the end of the bolus, and the open circle
disappears.
Easy bolus
The EASY BOLUS button allows a quick way to deliver a Normal bolus. You will preset the settings for
this feature in the EASY BOLUS OPTION screen in the BOLUS MENU. Your pump is sent from the factory with
the Easy Bolus feature set to off. If you want to use Easy Bolus, turn it on.
After you set up Easy Bolus, with each
press, you can increase the Normal bolus amount by a fixed amount,
called a step. Before you can deliver an Easy Bolus, you must set the amount in the EASY BOLUS ENTRY screen.
This amount equals the number of units of insulin for each step. The maximum number of steps can equal
up to your maximum bolus limit. When using vibrate mode, EASY BOLUS is limited to 20 steps or maximum
bolus, whichever comes first.
After you set your step amount, you can program your Easy Bolus. When you are in the HOME screen, each
time you press
the Easy Bolus amount increases by one step. You will hear a beep or feel a vibration for
each step increase. Each beep is a different tone. This makes it easy for you to count the beeps while you
are programming your Easy Bolus.
Easy bolus setup
1 Go to the EASY BOLUS OPTION screen.
Main > Bolus > Bolus Setup > Easy Bolus
Select On/Set, then press ACT.
NOTE: If you are using the remote control, the Easy Bolus must be set to on.
Optimizing pump therapy 101