Use the button to deliver a Normal bolus at any time except during another Normal bolus. A Normal bolus
will temporarily interrupt a Square Wave or Dual Wave bolus that is delivering. After the Normal bolus is
finished, the Square Wave or Dual Wave bolus delivery will resume.
NOTE: If you want to use the pump-to-meter link, make sure the meter option is on. Refer to the Meter
option section in this chapter for instructions.
1 If you want a correction bolus, check your blood glucose with your blood glucose meter and go to step 2.
If you want to bolus for food, go to step 2.
2 Press
on your pump, or go to the BOLUS MENU, select Use Bolus Wizard, and press ACT.
3 The ENTER BG screen will appear.
a. If you are not using the meter:
Enter your blood glucose value. Press ACT and continue to step 4. If you are not entering a blood
glucose and want to bolus for food, select the dashes in the ENTER BG screen. The Bolus Wizard
feature will calculate the insulin needed for your food entry without considering your blood glucose
level. Press ACT and continue to step 4.
b. If you are using the meter, you must program your bolus within 12 minutes of the pump receiving
the reading from the meter. If more than 12 minutes have passed, the reading will no longer be
available from the screen and you must enter your blood glucose manually.
Your pump checks if the entered BG is within your target range. Press ACT to accept the blood
glucose value. You can also change this blood glucose value, if necessary, then press ACT.
4 The ENTER FOOD screen appears.
a. If this is a food bolus, enter the food value you will eat, then press ACT.
b. If this is a correction bolus, select 0 (zero) as the value, then press ACT.
5 The ESTIMATE DETAILS screen appears. Review the information on this screen. If you need to make any
changes, press ESC to return to the ENTER BG screen (step 3) and make changes as necessary.
6 Press ACT in the ESTIMATE DETAILS screen. The SET BOLUS screen appears with the estimated bolus
amount flashing. Change the amount if desired. Press ACT to accept and start delivering your bolus.
NOTE: If you have BG Reminder turned On, the BG REMINDER DURATION screen displays. It allows you
to set the duration before you are reminded to check your blood glucose after a bolus. See the BG
Reminder section in the Basic programming chapter for information about this feature.
7 The BOLUS DELIVERY screen appears. The pump will beep or vibrate at the start and end of the bolus.
As the bolus is delivered, the screen shows the bolus type and amount until the total units have been
delivered. The screen then returns to the HOME screen.
Using the Bolus Wizard feature88