User guide
Pumps are sent from the factory with default BG Target range of 5.6–5.6 mmol/L (100–100 mg/dL).
1 Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Setup > Bolus Wizard Setup > Edit Settings
2 Select BG Target, then press ACT.
3 The TARGET RANGE 1 screen appears. The low end of your BG Target range flashes on the screen.
4 Set the BG Target, then press ACT.
5 The high end of your BG Target range flashes on the screen. Enter the BG Target, then press ACT.
The start time for your first BG Target is midnight and cannot be changed.
6a. If your BG Target is outside of 5.0–7.8 mmol/L (90–140 mg/dL), the pump screen displays a warning
that the values are acceptable but outside normal range. Press ESC to change your BG Target or press
ACT to set this range.
b. If your BG Target is within 5.0–7.8 mmol/L (90–140 mg/dL), the SET START TIME 2 screen appears.
The dashes under the screen name flash. The first BG Target range is now set.
If you do not need a second BG Target range, press ESC and skip to the next section. If you need to
set up another BG Target range, follow steps 7 through 11.
7 In the SET START TIME 2 screen enter the time of the day you want this BG Target range to become
active.
8 Press ACT. The TARGET RANGE 2 screen appears.
9 The low end of your BG Target range flashes on the screen. Set the BG Target, then press ACT.
10 The high end of your BG Target range flashes on the screen. Enter the BG Target, then press ACT.
11 The SET START TIME 3 screen appears. The second BG Target range is now set.
12 If you do not need to set up any more BG Target ranges, press ESC. If you need to set up more BG Target
ranges, repeat steps 7 through 11 above for each one.
You are now ready to set up the Active insulin time.
About active insulin
Active insulin is the bolus insulin that has already been delivered to your body, but has not yet been used.
The pump considers your active insulin time setting in determining any active insulin still in your body from
prior boluses. This may help prevent hypoglycemia caused by over-correcting for high blood glucose.
The pump shows the active insulin amount in the ESTIMATE DETAILS screen during the Bolus Wizard
programming steps, the STATUS and the SET BOLUS screens. However, the active insulin amount is calculated
differently in the ESTIMATE DETAILS screen and appears with an asterisk (*Active Insulin). The active insulin
amount calculated in the Bolus wizard steps includes the insulin that has already been delivered and insulin
that is going to be delivered by the active Square Bolus.
Using the Bolus Wizard feature84