Setting the BG units
You can select mmol/L or mg/dL as your Blood Glucose Unit (measurement type). You can also set these
units in the Sensor and Capture Event menus.
1 Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Setup > Bolus Wizard Setup > Edit Settings
2 Select BG Units, then press ACT.
3 The BLOOD GLUCOSE UNITS screen displays. Select mmol/L or mg/dL, then press ACT.
4 The EDIT SETTINGS screen shows the BG units you selected. You are now ready to set your insulin
sensitivity.
Insulin sensitivity
Your insulin sensitivity is the amount your blood glucose (BG) level is reduced by one unit of insulin. This
value is used to calculate a suggested insulin dose to correct a high BG. Because this sensitivity may vary
throughout the day, your pump lets you set up to eight sensitivity settings. Your healthcare professional may
only have you program one or two insulin sensitivities when you first start using the Bolus Wizard feature.
Record your settings in the Bolus Wizard feature settings table in this chapter.
Insulin sensitivity values are normally between 1.1–5.6 mmol/L/u (20–100 mg/dL/u). If your value is outside
this range, a warning message will appear on the screen.
1 Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Setup > Bolus Wizard Setup > Edit Settings
2 Select Sensitivity, then press ACT.
3 The INS SENSITIVITY 1 screen appears. The default sensitivity value flashes on the screen.
4 Set the value for your first insulin sensitivity setting, then press ACT.
The start time for your first insulin sensitivity is midnight and cannot be changed.
5 The SET START TIME 2 screen appears. The dashes under the screen name flash. The first insulin
sensitivity is now set.
If you do not need a second insulin sensitivity, press ESC and skip to the next section. If you need to set
up another insulin sensitivity, follow steps 6 through 9.
6 In the SET START TIME 2 screen enter the time of the day you want this insulin sensitivity to become
active.
7 Press ACT. The INS SENSITIVITY 2 screen appears.
8 The default sensitivity value flashes. Select the value for this insulin sensitivity.
Using the Bolus Wizard feature82