Technical data

TTR31, TTR35 The most important technical data
Endress+Hauser 31
8. 4 Software history and compatibility overview
The release number on the nameplate and in the Operating Instructions indicates the change
status of the device: XX.YY.ZZ (example 01.02.01).
Software history
9 The most important technical data
9. 1 Power supply
Supply voltage
DC voltage version 12...30 V DC
Current consumption
Without load < 60 mA, with reverse polarity protection
Power supply failure
Behaviour in case of overvoltage (> 30 V)
The device works continuously up to 34 V DC without any damage. No damage is caused to
the device in case of a short-term overvoltage up to 1 kV (as per IEC 61000-4-5). If the supply
voltage is exceeded, the properties specified are no longer guaranteed.
Behaviour in case of undervoltage
If the supply voltage drops below the minimum value, the device switches off (status as if not
supplied with power = switch open).
XX Change in the main version.
Compatibility no longer provided. Device and Operating Instructions change.
YY Change in functionality and operation.
Compatibility provided. Operating Instructions change.
ZZ Trouble-shooting and internal modifications.
Operating Instructions do not change.
Date Release no.
device
Changes in software Documentation
06.2004 01.00.00 KA174r/09/en
(51008032)
12.2004 01.01.00 New analog electronics BA201r/09/en/02.05
(51009833)
02.2005 01.02.00 Internal BA201r/09/en/02.05
(51009833)
02.2006 01.02.01 Parameter functional safety for the optional
analog output is not applicable
BA229r/09/en/03.06
(71025405)