Troubleshooting
The screen remains blank or there is no signal :
No power supply
Batteries are discharged
Contrast adjustment is incorrect
Press RESET for at least 10 seconds
REMARK: Temporarily remove the battery and adapter in
case RESET is not helping.
RMS readout is incorrect :
Make sure that at least 1 and preferably even 2
periods are displayed.
The battery is discharged.
No signal on the oscilloscope display :
Time/div setting is in the wrong position. Try 1ms
or choose auto-setup mode.
The unit is in the ‘hold’ position
Trigger function is set in the “once” position
The programmed trigger level is not reached
(choose “run” mode)
Y position is wrong or need callibration, see page 30
The input signal is too high, change the volt/div.
setting or choose auto-setup.
Incorrect frequency readout:
An incorrect time/div. setting has been chosen.
(start at 250ns/div)
Voltage readout does not correspond with the
actual value :
The measuring probe is in the X10 position
The battery is discharged.
The zero refrence is not set correctly for DC
measurements.
TROUBLESHOOTING