Manual
EVM1400_v2 PEREL 10
Resistance
Range Resolution Accuracy
400Ω 0.1Ω
± (1.2% of rdg + 2 digits)
4kΩ 1Ω
40kΩ 10Ω
400kΩ 100Ω
4MΩ 1kΩ
40MΩ 10kΩ ± (2.0% of rdg + 5 digits)
Open circuit voltage: 0.25V
Overload protection: 250VDC or 250VAC RMS
Capacitance
Range Precision Accuracy
4nF 1pF
± (3.0% of rdg + 3 digits)
40nF 10pF
400nF 0.1nF
4µF 1nF
40µF 10nF
200µF 100nF ± (8.0% of rdg + 10 digits)
Overload protection: 250VDC or 250VAC RMS
Frequency
Range Resolution Accuracy
9.999Hz 0.001Hz ± (2.0% of rdg + 5 digits)
99.99Hz 0.01Hz
± (1.5% of rdg + 5 digits) 999.9Hz 0.1Hz
9.999kHz 1Hz
99.99kHz 10Hz
± (2.0% of rdg + 5 digits)
199.9kHz 100Hz
> 200kHz as reference
Hz range
Measurement range: 0~200kHz
Input voltage range: 0.5VAC – 10VAC RMS
Overload protection: 250VDC or 250VAC RMS
V range
Measurement range: 0~40kHz
Input voltage range: 0.5VAC – 600VAC RMS
Input impedance: 10MΩ
Max. input voltage: 600VDC or 600VAC RMS
mA range
Measurement range: 0~40kHz
Max. input current: 400mA DC or AC RMS
Overload protection: F500mA/250V quick fuse