Failure Measure, failure rectification
Error messageError message
Error messageError message
Error message Increase the span between empty and full adjustment.
Adjustment span
too small
E017
Error messageError message
Error messageError message
Error message Instrument error
General An error in the sensor was found.
Instrument error
E034 Posible causes are:
PA status 12 - supply voltage not within tolerance
- defective oscillator
- damage to the electrode or to the insulation
- electrode short-circuited against vessel wall
• Separate the sensor from power supply and then reconnect (cold
start).
• If the failure message does not extinguish, check the sensor cable
and the power supply (correct voltage)
• If the error message does not extinguish, check the sensor for
visible damage.
If you cannot find any faults, please call our service department.