Technical data
762 Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
35 Status Reporting
The status register bits are described in more detail in the following tables:
• Table 48
• Table 46
• Table 53
• Table 54
• Table 56
• Table 51
The status registers picture above shows how the different status reporting data
structures work together. To make it possible for any of the Standard Event Status
Register bits to generate a summary bit, the bits must be enabled. These bits are
enabled by using the *ESE common command to set the corresponding bit in the
Standard Event Status Enable Register.
Wai t
Trig
Ru n
Arm
Reg
AER?
Run bit set if oscilloscope not stopped
:OPERation:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable (Mask) Register
10234567
+
OP ER T RGMAVESB
RQS /
MSS
TRG
Reg
TER?
Trigger Event Register
*SRE
*SRE?
Service Request Enable (Mask) Register
*STB?
Status Byte Register
Outp u t
Queue
SRQ
*ESR?
(Standard) Event Status Register
*ESE
*ESE?
(Standard) Event Status Enable (Mask) Register
PON URQ EXE DDE QYE RQL OPCCM E
OV L R
USR
MSG
+
OR
10234567
+
OR
1023456798101112131415
Service Request
OR
Wai t
Trig
Ru n
:OPERation[:EVENt]?
Operation Status Event Register
OV L R
11
5 3
From Overload
Event Registers
From Mask Test
Event Registers
MTE
MTE
9