Technical data

Obsolete and Discontinued Commands 33
Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 733
:MEASure:VDELta
(see page 792)
Query Syntax
:MEASure:VDELta?
The :MEASure:VDELta? query returns the voltage difference between vertical
marker 1 (Y1 cursor) and vertical marker 2 (Y2 cursor). No measurement is made
when the :MEASure:VDELta? query is received by the oscilloscope. The delta value
that is returned is the current value. This is the same value as the front-panel
cursors delta Y value.
VDELta = value at marker 2 - value at marker 1
Return Format
<value><NL>
<value> ::= delta V value in NR1 format
See Also "Introduction to :MARKer Commands" on page 318
"Introduction to :MEASure Commands" on page 340
":MARKer:Y1Position" on page 327
":MARKer:Y2Position" on page 328
":MARKer:YDELta" on page 329
":MEASure:TDELta" on page 727
":MEASure:TSTArt" on page 729
NOTE
The :MEASure:VDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:YDELta command (see page 329) instead.