Technical data
:TRIGger Commands 29
Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 603
:TRIGger[:EDGE]:LEVel
(see page 792)
Command Syntax
:TRIGger[:EDGE]:LEVel <level>
<level> ::= <level>[,<source>]
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<level> ::= ±(external range setting) in NR3 format
for external triggers
<level> ::= ±8 V for digital channels (MSO models)
<source> ::= {CHANnel<n> | EXTernal} for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | EXTernal}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :TRIGger[:EDGE]:LEVel command sets the trigger level voltage for the active
trigger source.
Query Syntax
:TRIGger[:EDGE]:LEVel? [<source>]
The :TRIGger[:EDGE]:LEVel? query returns the trigger level of the current trigger
source.
Return Format
<level><NL>
See Also • "Introduction to :TRIGger Commands" on page 589
• ":TRIGger[:EDGE]:SOURce" on page 606
• ":EXTernal:RANGe" on page 274
• ":POD<n>:THReshold" on page 415
• ":DIGital<d>:THReshold" on page 249
NOTE
If the optional source is specified and is not the active source, the level on the active source is
not affected and the active source is not changed.