Technical data

586 Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
28 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 792)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full-scale horizontal time in
seconds for the zoomed (delayed) window. The range is 10 times the current
zoomed view window seconds per division setting. The main sweep range
determines the range for this command. The maximum value is one half of the
:TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window timebase
range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 578
":TIMebase:RANGe" on page 581
":TIMebase:POSition" on page 580
":TIMebase:SCALe" on page 583