Technical data
:MTESt Commands 21
Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 383
:MTESt:AMASk:CREate
(see page 792)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask around the
current selected channel, using the tolerance parameters defined by the
:MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and :MTESt:AMASk:UNITs
commands. The mask only encompasses the portion of the waveform visible on
the display, so you must ensure that the waveform is acquired and displayed
consistently to obtain repeatable results.
The :MTESt:SOURce command selects the channel and should be set before using
this command.
See Also • "Introduction to :MTESt Commands" on page 379
• ":MTESt:AMASk:XDELta" on page 386
• ":MTESt:AMASk:YDELta" on page 387
• ":MTESt:AMASk:UNITs" on page 385
• ":MTESt:AMASk:SOURce" on page 384
• ":MTESt:SOURce" on page 409
Example Code • "Example Code" on page 379