Technical data
:LISTer Commands 18
Keysight InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 315
:LISTer:DISPlay
(see page 792)
Command Syntax
:LISTer:DISPlay <value>
<value> ::= {{OFF | 0} | {SBUS1 | ON | 1} | ALL}
The :LISTer:DISPlay command configures which of the serial buses to display in
the Lister, or whether the Lister is off. "ON" or "1" is the same as "SBUS1".
When set to "ALL", the decode information for different buses is interleaved in
time.
Serial bus decode must be on before it can be displayed in the Lister.
Query Syntax
:LISTer:DISPlay?
The :LISTer:DISPlay? query returns the Lister display setting.
Return Format
<value><NL>
<value> ::= {OFF | SBUS1 | ALL}
See Also • "Introduction to :LISTer Commands" on page 313
• ":SBUS<n>:DISPlay" on page 450
• ":LISTer:DATA" on page 314