Specifications

Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 77
Chapter 10: Setting the General tab preferences
Preference Description
View Blood Pressure Displays the red numeric blood pressure value
View Temperature Displays the blue numeric temperature value
3. Click OK.
The live data monitor panel displays the real time vital signs of the animal
based on the preferences you selected.
Live data monitor panel highlighting the real-time values for the selected traces
Related information
Physiological Enable preferences (page 75)
Physiological Alarm Levels (page 77)
Physiological Alarm Levels
Use the Physiological Alarm Levels preferences section to set the low and high
physiological data limits beyond which the system displays the pulsing red alarm
signal as shown in the following illustration.
You can specify the limits for ECG, respiration and temperature.
f To set the physiological data threshold levels:
1. From the Study Browser (page 49) click Prefs and then click the General tab.
2. In the Physiological Alarm Levels section:
a. If you want to activate the alarm for one of the data inputs, select the
appropriate check box.
b. Type your desired limit values in the Lower and Upper boxes.
3. Click OK.