Specifications

Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 71
Chapter 10: Setting the General tab preferences
Cine Loop Size preferences
Use the Cine Loop Size section to specify the amount of continuous image data
you want the system to keep in memory when you acquire a cine loop.
Cine Loop Size section displaying the default cine loop size values for each Mode that
supports cine loops
While you acquire data, the system's playback memory holds your most recent
image data in a buffer. The size of the buffer is determined by the Cine Loop Size
preference you specify.
Examples:
If you set your B-Mode cine loop size to 100 frames and you scan in B-Mode
for two minutes, when you press
Cine Store
or
Scan/Freeze
the system records
only the last 100 frames of image data that you acquired.
If you set your M-Mode cine loop size to 5 seconds and you scan in M-Mode
for two minutes, when you press
Cine Store
or
Scan/Freeze
the system records
only the last 5 seconds of image data that you acquired.
f To set the number of frames or seconds for a cine loop:
1. From the Study Browser (page 49) click Prefs and then click the General tab.
2. In the Cine Loop Size section, type a value in the appropriate box.
3. If you select Max for B-Mode, Contrast Mode or CF Doppler Modes, the
system sets the cine loop to acquire the maximum number of image frames
based on the current configuration of the system.
4. Click OK.
The system saves your preferences.