Specifications

Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 439
Appendix H: Technical support and user maintenance
3. Replace the ball and the housing ring.
f To disinfect the system:
Use Sporicidin wipes.
Maintaining the MicroScan transducer
The MicroScan transducer is the most delicate component of the Vevo 2100
Imaging System. Use care when handling the transducer. Proper handling
maintains the high quality performance of the transducer in addition to extending
the working lifetime of the transducer and the transducer.
WARNING: High levels of ultrasound energy can damage tissue. Do not
touch the transducer when acoustic power could be generated. Always
switch off the Vevo 2100 before attempting any cleaning or replacing the
transducer.
Cleaning the transducer
After every imaging session, gently wipe down the transducer with a soft cloth
and isopropyl alcohol or use Sporacidin wipes.
Storing the transducer
The transducer may be stored in the holder attached to the sides of the Vevo 2100
Imaging System.
Place the transducer into one of the holders with the nose pointing upward.
Always ensure that the cable is not twisted when storing the transducer.
Always use the provided case to transport the transducer from one site to
another.
f Follow these guidelines when you store the transducer in the
provided case:
Make sure that the transducer is clean and dry before you place it in the case.
Place the transducer in the case carefully to prevent kinking of the cable.
Avoid storing the transducer in areas of extreme temperatures or in direct
sunlight.
Store the transducer separately from other instruments to avoid inadvertent
damage.