Specifications

Rev 1.1 VisualSonics Vevo 2100 Imaging System Operator Manual 403
Appendix B: Troubleshooting
Problem Solution
Low frequency noise level in PW
Doppler acquisition is high
Increase the Wall Filter setting.
Noise appears in the image
Adjust the Sample Volume size and position such that it
includes tissue only.
3D-Mode
Problem Solution
Can’t initialize the motor
Ensure that the cable for the 3D motor stage is connected to
the rear panel.
Ensure that the motor is positioned such that there are no
objects obstructing the path of the transducer during
initialization.
Expected data is not acquired
Ensure the transducer is oriented correctly, with the transducer
arm of the transducer moving perpendicular to the direction of
travel of the 3D motor stage.
Ensure that the Range and Step Size settings are adequate for
acquiring the desired amount of data.
If two transducers are connected, ensure that the active
transducer is the one connected to the 3D motor stage.
Ensure that the transducer is tightly connected to the port on
the front of the cart.
Power Doppler Mode
Problem Solution
Color bands in the image
Enable Respiration Gating.
Adjust Wall Filter setting.
Adjust Scan Speed setting.
Adjust the Priority settings.
Respiration artifacts in the image
Enable Respiration Gating.
Adjust Wall Filter setting.
Adjust Sweep Speed setting.
Lack of sensitivity
Ensure the anatomy being studied is in the focal zone for the
transducer.