Specifications

366 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Appendix A: Measurement package protocols
Measurement package protocols
This appendix details the measurement and calculation definitions for each
measurement package that is available with the Vevo 2100 Imaging System.
In this appendix
Abdominal Measurement Package...................................................................................366
Cardiac Measurement Package .........................................................................................372
Embryology Measurement Package .................................................................................385
Ophthalmology Measurement Package ...........................................................................386
Vascular Measurement Package........................................................................................387
Abdominal Measurement Package
This section provides the measurements and calculations information for the protocols in the
Abdominal measurement package.
Liver protocol
Measurement definitions
Label Description Units Generic type Mode
Liver Sagg Sagittal length mm Linear B-Mode
Liver Trans Transverse length mm Linear B-Mode
Hepatic Vel Hepatic vein velocity mm/s Velocity PW Doppler
Hepatic Diam Hepatic vein diameter mm Linear B-Mode
Hepatic Diam Hepatic vein diameter mm Depth M-Mode
RHV Vel Right hepatic vein velocity mm/s Velocity PW Doppler
RHV Diam Right hepatic vein diameter mm Linear B-Mode
RHV Diam Right hepatic vein diameter mm Depth M-Mode
LHV Vel Left hepatic vein velocity mm/s Velocity PW Doppler
LHV Diam Left hepatic vein diameter mm Linear B-Mode
LHV Diam Left hepatic vein diameter mm Depth M-Mode
CHA Vel Common hepatic artery velocity mm/s Velocity PW Doppler
Appendix A