Specifications
276 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 41: Acquiring 3D-Mode images
a. Initializes the motor stage.
b. Confirms the initialization and prompts you to start the 3D slices
acquisition.
3. Click OK. The system returns to the base image acquisition Mode.
4. Activate the base imaging Mode for the type of 3D-Mode image you want to
acquire:
Press
B-Mode
to acquire a B-Mode only 3D image. The system begins to
acquire B-Mode image data.
Press
Power
to acquire Power Doppler Mode 3D image data over each B-
Mode image slice. The system begins to acquire Power Doppler Mode
image data.
5. Locate the object of interest and center it as closely as possible relative to the
transducer using the platform controls on the Vevo Imaging Station.
WARNING: Ensure that the lateral movement of the 3D motor stage cannot injure
the subject and damage the transducer.
6. Press
3D
.