Specifications
232 VisualSonics Vevo 2100 Imaging System Operator Manual Rev 1.1
Chapter 35: Acquiring M-Mode images
Update
Function 1: display control
Alternates the display from the dual view (B-Mode scout window on top, Mode image
window on the bottom) to the B-Mode image plus overlay so you can position your
sample gate more precisely.
To use this toggle control:
1. Press to view the dual view.
2. Press again to display the B-Mode window and overlay.
Function 2: right-click button
When the manual directs you to right-click, press
Update
.
SV/Gate
Push up to increase. Pull back to decrease.
In M-Mode: This control adjusts the size of the sample gate, measured in mm. The control
adjusts the distance of the vertical line between the two yellow calipers.
In the dual window view, the system displays the M-Mode sample gate image data.
Current data is on the right side, trailing data extends to the left.
Sweep Speed
Adjusts the cine loop playback speed parameter so that you can stretch out or compress
the cine loop data in the review window. Push up to increase the speed and compress the
cine loop image. Pull down to decrease the speed and expand the cine loop image.
When you are reviewing the cine loop you can also use the
Cine Loop Review
control to
adjust the sweep speed.
In M-Mode: Set the sweep speed parameter in a range from 200 Hz to 4000 Hz in
increments of 100 Hz. The system displays the updated values in the status bar in the
lower left area of the screen.
In cardiac applications you might want to decrease the M-Mode sweep speed so you can
view more wall movements over more cardiac cycles in the window, or increase the
speed so you can view more wall detail over one cycle.