Chapter 34: Analyzing B-Mode images
7. If you have selected Show Values and Labels in the Measurement tab of the
Preferences window and you want to move the measurement or move the
label, select either item and then drag and drop it.
8. Right-click the contour and select Replicate Forward or Replicate Reverse
and select the number of cycles. The system automatically traces the wall
forward or backward through the frames.
9. Press
Cine Store
to save the cine loop.
When you play the cine loop, the system displays the contour that represents the
systolic LV in green, and the diastolic LV in red.
Next step
Reporting your analysis results (page 184)
Related information
Analyzing image data (page 156)
Time Interval measurement
Time interval is measured in ms.
f To place a time interval measurement:
1. Click the time interval measurement button . The system highlights the
button until you complete your measurement.
2. In the physiology data trace window below the image mode data, click to
place the initial caliper.
3. Trackball to the location where you want to place your end caliper and then
click to place the caliper.
4. If you need to move the entire measurement, click on a measurement line and
then drag and drop it.
Related information
Complete procedure for adding a measurement (page 166)
Adding protocol measurements
Protocol measurements are labeled uniquely for a specific measurement protocol.