CALIBRATION & ADJUSTMENTS
5-27
10. To end dark current, press F1 (it defaults to AUTOLUMA).
11. Install the photometer and adjust Brightness Pots until the meter reads 6.0 ft-lamberts.
Adjust Autoluma pot VR1 on Logic board until Autoluminant reads 200 on Gentest menu.
12. Press F7 and verify that brightness reads greater than 8.00 ft-lamberts.
13. Press F3 to verify calibration.
14. Verify image alignment according to the procedure at the beginning of this chapter.